Xenon Arc Lamp Artificial Climate Aging Test Chamber
GZRM-XII1200

Overview of the Test Chamber
The xenon lamp artificial aging test chamber mainly uses a water-cooled long-arc xenon lamp as the light source. By adjusting temperature (sample rack temperature, working chamber temperature), relative humidity, rainfall method, etc., it simulates the comprehensive combination of light, heat and water under outdoor climatic test conditions to conduct artificial aging tests on the photochemical energy and mechanical properties of test materials. The main control parameters include light irradiance, working chamber temperature, sample rack temperature and relative humidity, which are automatically controlled through closed loops. This instrument adopts an advanced PLC controller, featuring advanced and stable control; various alarm messages are provided during the control process, ensuring good maintainability. The device complies with domestic xenon lamp aging test standards and international standards such as AATCC, ISO, DIN, and ASTM. It has a wide range of applications, including plastics, coatings, rubber, textiles, automobiles, building materials, etc. The control system of this instrument consists of: xenon lamp automatic control system; xenon lamp cooling system; temperature control system; humidity control system; and xenon lamp protection alarm system.
Technical Specifications
| Total Exposure Area | 11000cm2 | |
| Dimensions | 1080×1080×1080 (W×H×D in mm) | |
| Temperature Range | Working Chamber Temperature | 35℃~80℃ |
| Black Panel Temperature | 40℃~110℃ | |
| Black Standard Temperature | 40℃~120℃ | |
| Relative Humidity Range | Light Exposure | 10~75% |
| Dark Period | 10~95% | |
| Light Irradiance | Optional automatic control of wavelengths 340nm, 420nm or 300~400nm | |
0.20/㎡~1.39 W/㎡ @340nm 0.67W /㎡~3.11W/㎡ @420nm 26W /㎡~180 W/㎡ @(300~400nm) | ||
| Sample Rack Type | SL-3T type, single-layer structure | |
Main Advantages
|
|
|
|
|
|
|
|
|
|
|
|
Test chamber technical parameters
Laboratory environmental conditions
| Ambient Temperature | 23±2℃ |
| Ambient Relative Humidity | <80% |
Temperature
| Temperature Range | Working Chamber Temperature | 35℃~80℃ |
| Black Panel Temperature | 40℃~110℃ | |
| Black Standard Temperature | 40℃~120℃ | |
| Temperature Fluctuation | ±1℃ | |
Humidity
| Relative Humidity Range | Light Exposure | 10~75% |
| Dark Period | 10~95% | |
| Relative Humidity Fluctuation | ±4% | |
Light irradiance
| Light Irradiance | Optional automatic control of wavelengths 340nm, 420nm or 300~400nm |
0.20/㎡~1.39 W/㎡ @340nm 0.67W /㎡~3.11W/㎡ @420nm 26W /㎡~180 W/㎡ @(300~400nm) 25W /㎡~130 W/㎡ @(300~400nm) | |
| Light Irradiance Deviation | ±0.02W/m2 |
| High Irradiance Intensity Control | Can reach 2 suns or higher |
| Total Exposure Area | 11000cm2 |
| Standard Test Sample Holder | 110 pieces (150mm×70mm) or customized according to user requirements |
| Light Irradiance Calibration Function | Uses calibrated xenon lamps or irradiance meters Verified by national metrology departments |

Spectral Comparison Between the Light Source of GZRM-X Series Xenon Arc Lamp Artificial Climate Test Chamber and Sunlight
Test Methods and Procedures
| Test Methods and Program Capacity | Fixed: 12 international and national xenon lamp aging test standards |
| Editable: 4 programs, maximum 150 program segments can be set | |
| Simulation of Different Test Conditions | Using different internal and external filters: Boro/Boro, Quartz/Boro, Quartz/Quartz, Boro/Sodalime, etc. |
| Light/Dark Cycle | Automatically controlled, real-time monitored and displayed |
| Satisfied Test Standards | GB/T1865-1997 ;GB/T16422.2-2014 ;GB/T1189-91 GB/T8427-2008 GB/T8430 GB/T15102-2006 GB/T15104-2006 SAEJ1885 ;SAEJ19960 GB/T14049-2008 ISO 105B-06 ;ISO 105B-02 ;ISO 105B-04 ISO4892-2 ;ISO11341 AATCC16E ;AATCC169(1) JASOM346-93 GMW3414TM Ford Interior ASTM G155-1 ASTM G155-4 PV3929,PV3930,PV1303,PV1306,GMW14162, SMTC 5 400 003 |

Simulation of Different Test Conditions by Different Filters
Sample Requirements
Prohibited Items for This Test Equipment
Tests or storage of flammable, explosive, volatile substance samples, corrosive substance samples, biological samples, or samples with strong electromagnetic emission sources are prohibited.
Structural Features of the Test Chamber
Cabinet Body and Thermal Insulation Enclosur
| External Cabinet | 1250×1960×1460(W×H×D)mm |
| High-quality painted steel plate | |
| Internal Cabinet | 1080×10880×10880(W×H×D)mm |
| Stainless steel SUS316 | |
| Thermal Insulation Material | Rigid foam |
| Weight | 650kg |
Working Door
|
|
|
|
Sample Rack
|
|
|
|
Xenon Lamp Automatic Control System
| Control Method | Automatic control to ensure that xenon lamp irradiance does not attenuate with xenon lamp aging |
| Calibration | Uses calibrated xenon lamps or irradiance meters |
| System comes with calibration program | |
| Water Cooling Method of Water-Cooled Xenon Lamp | Automatically controls the temperature of cooling water |
| Imported water pump from Italy, all-stainless steel air cooler | |
| Environmentally friendly water-saving design | |
| Flow switch and water shortage protection | |
| Water-Cooled Xenon Lamp | 12KW, service life over 2000 hours, easy to disassemble |
Electrical Control System
| Operation Mode | Programmed operation |
| Setting Method | Chinese menu, touch screen input |
| Temperature and Humidity Control System | Centrifugal fan |
| Frequency converter for speed regulation | |
| Air damper regulator | |
| Ultrasonic humidifier | |
| Stainless steel heating wire: contactless SSR heating control method | |
| Air filter | |
| Spray System | Sample spray, sample rack spray |
| Programmed automatic control | |
| Water Tank Filling System | Automatic water filling |

Temperature and Humidity Control System
Protection System
| Main Power Supply and Branch Power Supply Switches | Siemens switches |
| Xenon Lamp System | Xenon lamp cooling water shortage protection |
| Xenon lamp cooling water over-temperature protection | |
| Xenon lamp irradiance low (high) alarm | |
| Xenon lamp cooling water level low protection | |
| Door safety protection | |
| Temperature Control System | Sample rack temperature over-temperature protection |
| Sample rack temperature low alarm | |
| Working chamber temperature over-temperature protection | |
| Blower motor abnormal protection | |
| Humidification System | Humidity display abnormal alarm |
| Others | Leakage, overcurrent and overload protection |
| Pure water resistivity abnormal alarm |
Touch Screen
| Siemens switches | ||
| Xenon Lamp System | Xenon lamp cooling water shortage protection | ||
| Xenon lamp cooling water over-temperature protection | |||
| Xenon lamp irradiance low (high) alarm | |||
| Xenon lamp cooling water level low protection | |||
| Door safety protection | |||
| Temperature Control System | Sample rack temperature over-temperature protection | ||
| Sample rack temperature low alarm | |||
| Working chamber temperature over-temperature protection | |||
| Blower motor abnormal protection | |||
| Humidification System | Humidity display abnormal alarm | ||
| Others | Leakage, overcurrent and overload protection | ||
| Pure water resistivity abnormal alarm |