Fluorescent UV Chamber
General Remarks
Scope of Delivery
The quotation includes a complete fluorescent UV chamber:
Fluorescent UV Chamber
Specimen Holders
Sensors
Commissioning in China
2. Technical Description of the Fluorescent UV Chamber
2.1 Irradiance Parameter
a)Wavelength
UVA 340nm, UVA 351nm, UVB 313nm, UVC 254nm.
b)Exposure Area
The Fluorescent UV Chamber provides two test chambers, located upper and bottom. Operate separately. Each Test chamber could hold 12 specimen holders (48 standard specimens).
c)Controlled Irradiance
Narrow band 340 nm (0.25-1.55 W/m2) or 351 nm (0.25-1.35 W/m2)or 313 nm (0.28-1.25 W/m2) or 254nm (0.25-1.35 W/m2) to meet test requirements.
(0.25-1.55 W/m2) (0.25-1.35 W/m2)
(0.28-1.25 W/m2) (0.25-1.35W/m2)
Irradiance Deviation
The Fluorescent UV Chamber provides following performance of irradiance deviation:
@340nm ±0.01W/m2
@351nm ±0.01W/m2
@313nm ±0.01W/m2
@254nm ±0.01W/m2
2.2 Technical System Description
a)Illuminant
UV fluorescent tube is used in the weathering testing system, which is the closest simulation of UV of solar radiation.
Bulb Type: UVA/UVB/UVC Fluorescent Tube
Power: 40 W
Lifetime: 4,000 hrs
Wavelength: UVA 340nm or 351nm, UVB 313nm, UVC 254nm
b)Specimen Holder
Provide 24pcs of adjustable specimen holders (75mm x 300mm) for 48pcs standard specimens (75mm x 150mm). Specimen holders could be customized.
Distance between center of tube to surface of specimen: 50mm ± 3mm
c)Condensation Control
The heater provides control of condensation and enables automatic control and display of condensation inside the chamber.
Humidity range during condensation: ≥ 95% RH
d)Temperature Control
Temperature inside of the chamber, black panel thermometer (BPT) temperature or black standard thermometer (BST) temperature are controlled or monitored consistently and simultaneously.
Automatic air circulation system for recirculating chamber air.
Temperature range inside of the chamber (maximum ambient temperature of 35°C): 40°C to 90°C
Temperature uniformity inside of the chamber: ≤ 2°C
Temperature range of BPT: 40°C to 110°C
Temperature range of BST: 40°C to 120°C
All temperature fluctuations: ± 1°C
e)Specimen Spray
Nozzles inside of the chamber provide uniform spraying of specimens with deionized water.
The specimen spray applies water to the exposed surface of the specimen which simulates rain to induce temperature shock and erosion effects.
f)Control System
The control system of the Fluorescent UV Chamber is a PLC, Type Siemens.
The system will be delivered with PLC program and a touch screen control panel color display.
g)Software Functionality
Setting and closed loop control of:
- Irradiance
- BPT/BST Temperature
- Condensation
- Spray
- Chamber Temperature
Display of Test Parameters and Progress
Monitor and Alarm function of water level and system fault.
14 factory pre-programmed international standard test methods.
4 custom test programs, which could be fully customized, such as step by step adjustment of irradiance, chamber temperature, BST temperature, condensation and spray to achieve customized testing procedure or cycles. All existing test programs can be saved, loaded and edited for custom applications.
Multi-Language operation system (English, German, Chinese)
h)Irradiance Calibration
The Fluorescent UV Chamber equipped with an automatic program for irradiance calibration, with could be used with a calibration radiometer.
i)Testing Standards
The Fluorescent UV Chamber meets or exceeds the following industry standards:
- SAE J2020
- PREN 1062-4
- ISO4892-3
- ASTM G53
- ASTM G154, EN534
- DIN74324.1
- ISO11507
3. Technical Data of the Fluorescent UV Chamber
Lamp: Fluorescent UV
Bulb Type: UVA 340nm or 351nm, UVB 313nm, UVC 254nm
Bulb Power: 40 W
Lifetime Bulb: 4,000 h
Irradiance: 0.25-1.55 W/m2 @ 340 nm
0.25-1.35 W/m2 @ 351 nm
0.28-1.25 W/m2 @ 313 nm
0.25-1.25 W/m2 @ 254 nm
Irradiance Deviation: ±0.01W/m2 @ 340 nm
±0.01W/m2 @ 351 nm
±0.01W/m2 @ 313 nm
±0.01W/m2 @ 254 nm
Specimen Rack: Sigle tier structure
Specimen Holder: 24pcs, 75 mm x 300 mm
Humidity: during condensation: ≥ 95% RH
Temperature Range: Inside of the chamber: 40°C to 90°C (ambient 35°C)
Uniformity inside of the chamber: ≤ 2°C
BPT: 40°C to 110°C
BST: 40°C to 120°C
All fluctuations: ± 1°C
Spray: Specimen spray
rradiance Calibration: with calibration radiometer
Size: 560 mm x 820 mm x 1880 mm (D x W x H)
Installation type: 3Ph / N / PE, 380 ± 10%VAC, 50 ± 1 Hz
Installation load: 3 kW
Ambient Temperature: 10°C to 35°CAmbient Humidity: 10% RH to 85% RH
Water Supply: Deionized water
Pressure: 193kPa (28PSI)
Temperature: ≤ 18.5°C
Flow rate: 7.0 L/min
Keywords