Xenon Arc Lamp Artificial Climate Aging Test Chamber
GZRM-X65

Technical Specifications
| Total Exposure Area | 6500cm2 | |
| Dimensions | Inner Chamber: 780×790×780 (W×H×D in mm) | |
| Outer Dimensions: 1270×1950×1100 (W×H×D in mm) | ||
| Temperature Range | Working Chamber Temperature | 35℃~85℃ |
| Black Panel Temperature | 40℃~110℃ | |
| Black Standard Temperature | 40℃~120℃ | |
| Relative Humidity Range | Light Exposure | 10~75% |
| Dark Period | 10~95% | |
| Light Irradiance | Optional Wavelengths: 340nm, 420nm or 300~400nm, with automatic control | |
0.28 /㎡~1.6 W/㎡ @340nm 0.6W /㎡~3.0W/㎡ @420nm 30W /㎡~180 W/㎡ @(300~400nm) | ||
| Sample Rack Type | Type AH, Three-layer Structure | |
| Test Sample Holder | Standard Rack: 65 pieces (150×70mm) or customized according to user requirements | |
Main Advantages
|
|
|
|
|
|
|
|
|
|
|
Test chamber technical parameters
Laboratory environmental conditions
| Ambient Temperature | 23±2℃ |
| Ambient Relative Humidity | <80% |
Temperature
| Temperature Range | Working Chamber Temperature | 35℃~85℃ |
| Black Panel Temperature | 40℃~110℃ | |
| Black Standard Temperature | 40℃~120℃ | |
| Temperature Fluctuation | ±2℃ | |
Humidity
| Relative Humidity Range | Light Exposure | 10~75% |
| Dark Period | 10~95% | |
| Relative Humidity Fluctuation | ±4% | |
Light Irradiance
| Light Irradiance | Optional Wavelengths: 340nm, 420nm or 300~400nm, with automatic control |
0.28/㎡~1.6 W/㎡ @340nm 0.6W /㎡~3.0W/㎡ @420nm 30W /㎡~180 W/㎡ @(300~400nm) | |
| Light Irradiance Deviation | ±0.02W/m2@340nm\@420nm; ±2W/m2@300-400nm; |
| High Irradiance Intensity Control | Up to 2 suns or higher |
| Total Exposure Area | 6500cm2(1008 in2) |
| Standard Test Sample Holder | 65 pieces (150mm×70mm) or customized according to user requirements |
| Light Irradiance Calibration Function | Calibrated with calibration xenon lamp or irradiance calibration meter, certified by national metrology authorities |

Spectral Comparison Between the Light Source of GZRM-X Series Xenon Arc Lamp Artificial Climate Test Chamber and Sunlight
Test Methods and Procedures
| Test Methods and Program Capacity | Fixed: 12 international and national xenon lamp aging test standards |
| Editable: 4 programs, maximum 150 program segments can be set | |
| Simulation of Different Test Conditions | Using different internal and external filters: Boro/Boro, Quartz/Boro, Quartz/Quartz, Boro/Sodalime, etc. |
| Light/Dark Cycle | Automatically controlled, real-time monitored and displayed |
| Satisfied Test Standards | GB/T1865-2009 GB/T16422.2-2014 GB/T1189-1991 GB/T12527-2008 GB/T14049-2008 GB/T8430-1998 GB/T2423.24-2013 SAE_J_2412_2004-旧1885 ;SAE J2527-旧1960 ISO 105B-06 ;ISO 105B-02 ;ISO 105B-04 ISO4892-2 ;ISO11341 AATCC16E ;AATCC169(1) JASOM346-93 IEC 68-2-18 MS 200-42-2009 E_M0135_2001-N GMW3414TM Ford Interior ASTM G155-1 ASTM G155-4 ASTM G155-5 ASTM G154 ASTM G151 ASTM D5423 |

Simulation of Different Test Conditions by Different Filters
Sample Rack
|
|
|
|

Rotary Drum Type Test Rack
Main Control Systems
Xenon Lamp Automatic Control System
| Control Method | Automatic control to ensure that xenon lamp irradiance does not attenuate with xenon lamp aging |
| Calibration | Calibrated with calibration xenon lamp |
| System comes with built-in calibration program | |
| Water Cooling Method for Water-Cooled Xenon Lamp | Automatic control of cooling water temperature |
| Imported high-quality water pump and full stainless steel cooling fan | |
| Environmentally friendly water-saving design | |
| Flow switch low water protection function | |
| Water-Cooled Xenon Lamp (American ATLAS) | 6.5KW, service life over 2000 hours, easy to disassemble |

Light Irradiance Calibration Function
This product adopts the automatic calibration function of standard xenon lamp. Every 400 hours, an alarm message will remind the user. It ensures the accuracy of light irradiance, a key indicator during the xenon lamp aging test, and the standard xenon lamp has been tested by the National Institute of Metrology, China, an authoritative national institution.
Protection System
| Main Power Supply and Branch Power Switch | Siemens Switch |
| Xenon Lamp System | Xenon lamp cooling low water protection |
| Xenon lamp cooling water over-temperature protection | |
| Xenon lamp irradiance low (high) alarm | |
| Xenon lamp cooling water level low protection | |
| Door safety protection | |
| Temperature Control System | Sample rack temperature over-temperature protection |
| Sample rack temperature low alarm | |
| Working chamber temperature over-temperature protection | |
| Blower motor abnormal protection | |
| Humidification System | Humidity display abnormal alarm |
| Others | Leakage, overcurrent and overload protection |
| Pure water resistivity abnormal alarm |
Instrument Acceptance and Metrological Traceability
- Acceptance Outline:
Xenon lamp aging test chamber
Chamber Temperature
Black Panel/Black Standard Temperature
Relative Humidity
Light Irradiance
Appearance and Accessories
Local Metrology Institute
Local Metrology Institute
National Institute of Metrology, China
Local Metrology Institute
Light Irradiance Calibration Standard:

2、Parameter Traceability
Traceability: Measurement results can be traced to international or national standards.
Traceability Diagram
International Bureau of Weights and Measures (BIPM)/Reference Materials
National Bureau of Metrology/Reference Materials
Temperature (Sample Rack Temperature/Working Chamber Temperature)
Relative Humidity
Light Irradiance (@340nm;@420nm,300-400nm)
Calibration Xenon Lamp
Irradiance Calibration Meter


Keywords
Xenon Arc Lamp Artificial Climate Aging Test Chamber GZRM-X65
Contact Us